4D Vessel Reconstruction for Benchtop Thrombectomy Analysis
- URL: http://arxiv.org/abs/2604.06671v1
- Date: Wed, 08 Apr 2026 04:45:17 GMT
- Title: 4D Vessel Reconstruction for Benchtop Thrombectomy Analysis
- Abstract summary: Mechanical thrombectomy can cause vessel deformation and procedure-related injury.<n>Time-resolved, full-field 3D vessel-motion measurements remain limited.<n>We developed a nine-camera, low-cost multi-view workflow for benchtop thrombectomy.
- Score: 0.0
- License: http://arxiv.org/licenses/nonexclusive-distrib/1.0/
- Abstract: Introduction: Mechanical thrombectomy can cause vessel deformation and procedure-related injury. Benchtop models are widely used for device testing, but time-resolved, full-field 3D vessel-motion measurements remain limited. Methods: We developed a nine-camera, low-cost multi-view workflow for benchtop thrombectomy in silicone middle cerebral artery phantoms (2160p, 20 fps). Multi-view videos were calibrated, segmented, and reconstructed with 4D Gaussian Splatting. Reconstructed point clouds were converted to fixed-connectivity edge graphs for region-of-interest (ROI) displacement tracking and a relative surface-based stress proxy. Stress-proxy values were derived from edge stretch using a Neo-Hookean mapping and reported as comparative surface metrics. A synthetic Blender pipeline with known deformation provided geometric and temporal validation. Results: In synthetic bulk translation, the stress proxy remained near zero for most edges (median $\approx$ 0 MPa; 90th percentile 0.028 MPa), with sparse outliers. In synthetic pulling (1-5 mm), reconstruction showed close geometric and temporal agreement with ground truth, with symmetric Chamfer distance of 1.714-1.815 mm and precision of 0.964-0.972 at $τ= 1$ mm. In preliminary benchtop comparative trials (one trial per condition), cervical aspiration catheter placement showed higher max-median ROI displacement and stress-proxy values than internal carotid artery terminus placement. Conclusion: The proposed protocol provides standardized, time-resolved surface kinematics and comparative relative displacement and stress proxy measurements for thrombectomy benchtop studies. The framework supports condition-to-condition comparisons and methods validation, while remaining distinct from absolute wall-stress estimation. Implementation code and example data are available at https://ethanuser.github.io/vessel4D
Related papers
- THA-Flow Generative Model: Prosthesis Geometry Prediction from Preoperative CT [7.351430886602578]
We present THA-Flow, a conditional flow-matching model that generates three-dimensional prosthesis geometry directly from preoperative CT.<n>Separate AutoencoderKL models compress preoperative bone anatomy and prosthesis geometry, while a three-dimensional UNet learns a rectified flow from Gaussian noise to the prosthesis latent space.
arXiv Detail & Related papers (2026-08-26T14:22:06Z) - A Multidimensional Data-Driven Hybrid Transformer Framework for Non-invasive Continuous Blood Pressure Prediction [28.58502100288247]
We propose a hybrid Transformer framework to estimate diastolic and systolic BP from ECG/III feature sequences.<n>Rather than raw waveforms, the framework models 10-step sequences of physiological descriptors and two demographic cos.
arXiv Detail & Related papers (2026-08-24T14:02:10Z) - Clinically Aligned Geometry Constraints for Robust IVUS Vessel Boundary Segmentation [10.359133930200992]
Intravascular ultrasound (IVUS) segmentation is important for quantitative coronary plaque burden assessment.<n>Standard methods prioritising overlap scores often suffer from boundary drift and topology errors, leading to inaccurate clinical measurements.<n>We present GeoCat, a geometry-consistent network that processes 5-frame IVUS clips using dual Cartesian-polar encoders with cross-domain attention and temporal fusion.<n>We evaluate performance using both segmentation accuracy and plaque-relevant clinical metrics, including Dice/IoU, boundary measures(95HD), ASSD, topology violation rate, and clinical geometry errors (dmax/dmin, angles, and areas)
arXiv Detail & Related papers (2026-06-17T06:02:01Z) - Anatomical Token Uncertainty for Transformer-Guided Active MRI Acquisition [27.825869252861335]
Full data acquisition in MRI is inherently slow, which limits clinical throughput and increases patient discomfort.<n>CSMRI seeks to accelerate acquisition by reconstructing images from under-sampled kspace data.
arXiv Detail & Related papers (2026-03-23T10:49:23Z) - Virtual Intraoperative CT (viCT): Sequential Anatomic Updates for Modeling Tissue Resection Throughout Endoscopic Sinus Surgery [0.8060701341619877]
We present Virtual Intraoperative CT (viCT), a method for sequentially updating preoperative CT throughout endoscopic sinus surgery (ESS)<n>viCT uses 3D reconstructions from monocular endoscopic video to enable visualization of evolving anatomy in CT format.<n>Results: viCT updates show agreement with ground-truth anatomy across surgical stages, with submillimeter mean surface errors.
arXiv Detail & Related papers (2026-03-07T00:23:05Z) - Leveraging Synthetic Priors for Monocular Depth Estimation in Specular Surgical Environments [0.0]
Existing self-supervised methods often suffer from boundary collapse on thin surgical tools and transparent surfaces.<n>In this work, we address this by leveraging the high-fidelity synthetic priors of the Depth Anything V2 architecture.<n>Our approach establishes a new state-of-the-art, achieving an accuracy ( 1.25) of 98.1% and reducing Squared Relative Error by over 17% compared to established baselines.
arXiv Detail & Related papers (2025-12-29T17:29:42Z) - End-to-end Cortical Surface Reconstruction from Clinical Magnetic Resonance Images [2.920414237330382]
We train the first neural network for explicit estimation of cortical surfaces from scans of any contrast and resolution.<n>Our method deforms a template mesh to the white matter (WM) surface, which guarantees topological correctness.<n>We show a approximately 50 % reduction in cortical thickness error (from 0.50 to 0.24 mm) with respect to recon-all-clinical (RAC) and better recovery of the aging-related cortical thinning patterns detected by FreeSurfer on high-resolution T1w scans.
arXiv Detail & Related papers (2025-05-20T07:18:58Z) - Monitoring morphometric drift in lifelong learning segmentation of the spinal cord [27.329843979159932]
We present a spinal cord segmentation model trained on a multisite $(n=75)$ dataset, including 9 different MRI contrasts and several spinal cord pathologies.<n>We also introduce a lifelong learning framework to automatically monitor the morphometric drift as the model is updated using additional datasets.
arXiv Detail & Related papers (2025-05-02T16:04:00Z) - Time-Varying Coronary Artery Deformation: A Dynamic Skinning Framework for Surgical Training [33.92599418560439]
This study proposes a novel dynamic modeling framework for coronary arteries using skeletal skinning weights.<n>It aims to achieve precise control over vessel deformation while maintaining real-time performance for surgical simulation applications.
arXiv Detail & Related papers (2025-03-04T02:51:37Z) - Registration of Longitudinal Spine CTs for Monitoring Lesion Growth [32.449176315776114]
We present a novel method to automatically align longitudinal spine CTs.
Our method follows a two-step pipeline where vertebrae are first automatically localized, labeled and 3D surfaces are generated.
We tested our approach on 37 vertebrae, from 5 patients, with baseline CTs and 3, 6, and 12 months follow-ups leading to 111 registrations.
arXiv Detail & Related papers (2024-02-14T17:43:50Z) - Spatial-Temporal Transformer based Video Compression Framework [44.723459144708286]
We propose a novel Spatial-Temporal Transformer based Video Compression (STT-VC) framework.
It contains a Relaxed Deformable Transformer (RDT) with Uformer based offsets estimation for motion estimation and compensation, a Multi-Granularity Prediction (MGP) module based on multi-reference frames for prediction refinement, and a Spatial Feature Distribution prior based Transformer (SFD-T) for efficient temporal-spatial joint residual compression.
Experimental results demonstrate that our method achieves the best result with 13.5% BD-Rate saving over VTM.
arXiv Detail & Related papers (2023-09-21T09:23:13Z) - Joint Rigid Motion Correction and Sparse-View CT via Self-Calibrating
Neural Field [37.86878619100209]
NeRF has widely received attention in Sparse-View (SV) CT reconstruction problems as a self-supervised deep learning framework.
Existing NeRF-based SVCT methods strictly suppose there is completely no relative motion during the CT acquisition.
This work proposes a self-calibrating neural field that recovers the artifacts-free image from the rigid motion-corrupted SV measurement.
arXiv Detail & Related papers (2022-10-23T13:55:07Z) - Cephalometric Landmark Regression with Convolutional Neural Networks on
3D Computed Tomography Data [68.8204255655161]
Cephalometric analysis performed on lateral radiographs doesn't fully exploit the structure of 3D objects due to projection onto the lateral plane.
We present a series of experiments with state of the art 3D convolutional neural network (CNN) based methods for keypoint regression.
For the first time, we extensively evaluate the described methods and demonstrate their effectiveness in the estimation of Frankfort Horizontal and cephalometric points locations.
arXiv Detail & Related papers (2020-07-20T12:45:38Z) - Appearance Learning for Image-based Motion Estimation in Tomography [60.980769164955454]
In tomographic imaging, anatomical structures are reconstructed by applying a pseudo-inverse forward model to acquired signals.
Patient motion corrupts the geometry alignment in the reconstruction process resulting in motion artifacts.
We propose an appearance learning approach recognizing the structures of rigid motion independently from the scanned object.
arXiv Detail & Related papers (2020-06-18T09:49:11Z) - Exploring Partial Intrinsic and Extrinsic Symmetry in 3D Medical Imaging [39.5958976939981]
We present a novel methodology to detect imperfect bilateral symmetry in CT of human anatomy.
The structurally symmetric nature of the pelvic bone is explored and is used to provide interventional image augmentation for treatment of unilateral fractures in patients with traumatic injuries.
arXiv Detail & Related papers (2020-03-04T19:08:55Z)
This list is automatically generated from the titles and abstracts of the papers in this site.
This site does not guarantee the quality of this site (including all information) and is not responsible for any consequences.