Traceable In Situ Microwave Power Measurement at the Cryogenic Device Plane in a Dilution Refrigerator
- URL: http://arxiv.org/abs/2607.12751v1
- Date: Tue, 14 Jul 2026 13:18:33 GMT
- Title: Traceable In Situ Microwave Power Measurement at the Cryogenic Device Plane in a Dilution Refrigerator
- Abstract summary: Accurate knowledge of the microwave power delivered to a cryogenic device under test is essential for the characterization and operation of superconducting quantum circuits.<n>This paper presents an in situ measurement method for RF power at the cryogenic device plane.
- Score: 32.63669289843474
- License: http://creativecommons.org/licenses/by-nc-nd/4.0/
- Abstract: Accurate knowledge of the microwave power delivered to a cryogenic device under test (DUT) is essential for the characterization and operation of superconducting quantum circuits. However, this information is difficult to obtain inside dilution refrigerators because of distributed attenuation, impedance mismatch, switch-path repeatability, and temperature-dependent microwave components. This paper presents an in situ measurement method for RF power at the cryogenic device plane. The method uses a custom variable temperature stage (VTS) as a cryogenic thermal-transfer element. The TVS is alternately heated by a four-wire DC heater and by microwave power dissipated in a 20 dB pass-through attenuator. By fitting the thermal transients and comparing the corresponding steady-state temperatures, the absorbed microwave power is inferred from a directly measured DC electrical power through an AC/DC substitution procedure. The finite reflection and transmission of the attenuator are then accounted for by cryogenic two-port scattering-parameter measurements based on a switch-assisted Short--Open--Load--Reciprocal calibration, so that the result is referred to the DUT reference plane. The system is demonstrated in a dilution refrigerator with powers between -43 and -58 dBm at the DUT input plane. The demonstrated relative standard uncertainty ranges from about 2% at -43.9 dBm to about 40% at -57.6 dBm. The proposed approach combines thermal RF power transfer, cryogenic S-parameter correction, and uncertainty evaluation in a measurement architecture compatible with quantum-device experiments, providing a practical route toward traceable microwave-power calibration at millikelvin stages.
Related papers
- Measurement and Calibration Approaches for Full Two-Port Scattering Parameters at mK Temperatures [32.73124984242397]
The Istituto Nazionale di Ricerca Metrologica (INRiM, Italy), exploits the Short-Open-Load-Reciprocal technique.<n>The system operates at temperatures down to the mK range, in the band 4-12 GHz in coaxial line.
arXiv Detail & Related papers (2025-05-26T12:45:30Z) - Thermalization of a flexible microwave stripline measured by a superconducting qubit [0.0]
Flexible microwave striplines offer the tantalyzing perspective of increasing the cabling density by an order of magnitude without thermally overloading the cryostat.
We use a superconducting quantum circuit to test the thermalization of input flex cables with integrated $60,$dB of attenuation distributed at various temperature stages.
We confirm that the qubit reaches an effective temperature of $26.4,$mK, close to the base temperature of the cryostat, practically the same as when using a conventional semi-rigid coaxial cable setup.
arXiv Detail & Related papers (2024-10-01T20:26:49Z) - Overcoming the Thermal-Noise Limit of Microwave Measurements by Pre-cooling with an Active Cold Load [0.34530027457862006]
We introduce a method, which we here name active pre-cooling'' (APC), for removing, just prior to performing a measurement, a large fraction of the deleterious thermal photons that would otherwise occupy the electromagnetic modes of a microwave cavity.<n>We report a room-temperature bench-top demonstration of the method, where this load takes the form of the input of a commercial low-noise amplifier (LNA)
arXiv Detail & Related papers (2024-08-09T22:43:29Z) - Millikelvin measurements of permittivity and loss tangent of lithium
niobate [50.591267188664666]
Lithium niobate is an electro-optic material with many applications in microwave signal processing, communication, quantum sensing, and quantum computing.
We present findings on evaluating the complex electromagnetic permittivity of lithium niobate at millikelvin temperatures.
arXiv Detail & Related papers (2023-02-24T22:05:42Z) - Precision measurement of the microwave dielectric loss of sapphire in
the quantum regime with parts-per-billion sensitivity [50.591267188664666]
Dielectric loss is known to limit state-of-the-art superconducting qubit lifetimes.
Recent experiments imply upper bounds on bulk dielectric loss tangents on the order of $100$ parts-per-billion.
We have devised a measurement method capable of separating and resolving bulk dielectric loss with a sensitivity at the level of $5$ parts per billion.
arXiv Detail & Related papers (2022-06-29T00:14:11Z) - Measurement of the Low-temperature Loss Tangent of High-resistivity
Silicon with a High Q-factor Superconducting Resonator [58.720142291102135]
We present the direct loss tangent measurement of a high-resist intrinsicivity (100) silicon wafer in the temperature range from 70 mK to 1 K.
The measurement was performed using a technique that takes advantage of a high quality factor superconducting niobium resonator.
arXiv Detail & Related papers (2021-08-19T20:13:07Z) - Cryogenic single-port calibration for superconducting microwave
resonator measurements [0.0]
In this paper, we demonstrate a data-based, single-port calibration using commercial microwave standards and a vector network analyzer (VNA)
We cross reference our data-based, single-port calibration and reflection measurement with over-coupled 2D- and 3D-resonators against well established two-port techniques corroborating the validity of our method.
arXiv Detail & Related papers (2021-03-02T06:24:51Z) - Adiabatic Sensing Technique for Optimal Temperature Estimation using
Trapped Ions [64.31011847952006]
We propose an adiabatic method for optimal phonon temperature estimation using trapped ions.
The relevant information of the phonon thermal distributions can be transferred to the collective spin-degree of freedom.
We show that each of the thermal state probabilities is adiabatically mapped onto the respective collective spin-excitation configuration.
arXiv Detail & Related papers (2020-12-16T12:58:08Z) - Real-time estimation of the optically detected magnetic resonance shift
in diamond quantum thermometry [47.50219326456544]
We investigate the real-time estimation protocols for the frequency shift of optically detected magnetic resonance (ODMR) of nitrogen-vacancy centers in nanodiamonds (NDs)
Efficiently integrating multipoint ODMR measurements and ND particle tracking into fluorescence microscopy has recently demonstrated stable monitoring of the temperature inside living animals.
arXiv Detail & Related papers (2020-06-12T01:44:35Z) - Microwave electrometry via electromagnetically induced absorption in
cold Rydberg atoms [9.897318014645432]
Direct International System of Units (SI)-traceable and self-calibrated method for measuring a microwave electric field strength based on electromagnetically induced absorption (EIA) in cold Rydberg atoms.
A narrower linewidth of cold Rydberg EIA enables us to realize a direct SI-traceable microwave-electric-field measurement as small as $sim$100.
arXiv Detail & Related papers (2020-02-03T16:04:06Z)
This list is automatically generated from the titles and abstracts of the papers in this site.
This site does not guarantee the quality of this site (including all information) and is not responsible for any consequences.